Characterization of a mini membrane inlet mass spectrometer for on-site detection of contaminants in both aqueous and liquid organic samples

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 20, 期 9, 页码 1441-1446

出版社

WILEY
DOI: 10.1002/rcm.2466

关键词

-

向作者/读者索取更多资源

A mini membrane inlet mass spectrometer (mini-MIMS) of a total weight of 12 kg was constructed using a miniature Multipole mass spectrometer, a small vacuum system and a flexible flat sheet membrane inlet, where the exposed membrane area can be changed by a factor of 80. The variable membrane area together with the possibility of operating the Multipole at pressures up to 1 x 10(-3) Torr made it possible to test the system with three microporous membranes (cellulose, polyether sulfone and polypropylene) normally not compatible with standard electron ionization MIMS systems and a standard non-porous polyclimethylsiloxane membrane. We found that the hydrophilic cellulose and polyether sulfone membranes had selectivity characteristics opposite to those of the standard silicone membrane. They demonstrated preferential detection of hydrophilic compounds in hydrophobic organic solvents, whereas the silicone membrane preferentially detects hydrophobic organic compounds in aqueous solution. Using the cellulose membrane, organic contaminants and water could be detected in organic solvents at 10-100 ppm levels by weight, the relative high detection limits primarily caused by interference from a high chemical background from the solvent. When being used with the standard silicone membrane the mini-MIMS behaved just like most standard MIMS systems with detection limits of volatile organic compounds in water at concentratiorts just below 1 ppm. The hydrophobic microporous polypropylene membrane was not found to be useful with the mini-MIMS. Copyright (c) 2006 John Wiley & Sons, Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据