DSC and high resolution X-ray diffraction coupling

期刊

JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY
卷 85, 期 1, 页码 219-224

出版社

SPRINGER
DOI: 10.1007/s10973-005-7351-y

关键词

lipid structures; mesophase and biomaterial studies; polymorphism; small angle X-ray scattering; triacylglycerol

向作者/读者索取更多资源

Coupling of time-resolved synchrotron X-ray diffraction at both small and wide angles with differential scanning calorimetry is a new technique that allows simultaneous characterization of thermal and structural properties of a sample. The apparatus, called Microcalix, works between -30 and +230 degrees C at scanning rates comprised between 0.01 and 20 degrees C min(-1) with a high sensitivity in both measurements using a single sample of small volume (from about I to 20 mu L). The last version of the instrument is designed for laboratory bench and conventional source but preferably with rotating anode or multilayered mirrors. Measurements under low pressure or under shear as well as recordings of isothermal evolution are also possible. The example of the study of polymorphism of a monounsaturated triglyceride (PPO) will be presented as an application.