4.4 Article

Rapid and sensitive determination of donepezil in human plasma by liquid chromatography/tandem mass spectrometry: application to a pharmacokinetic study

期刊

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
卷 20, 期 21, 页码 3193-3198

出版社

JOHN WILEY & SONS LTD
DOI: 10.1002/rcm.2718

关键词

-

向作者/读者索取更多资源

A liquid chromatography/tandem mass spectrometry (LC/MS/MS) method was developed and validated for the determination of donepezil in human plasma samples. Diphenhydramine was used as the internal standard. The collision-induced transition m/z 380 -> 91 was used to analyze donepezil in selected reaction monitoring mode. The signal intensity of the m/z 380 -> 91 transition was found to relate linearly with donepezil concentrations in plasma from 0.1-20.0 ng/mL. The lower limit of quantification of the LC/MS/MS method was 0.1 ng/mL. The intra- and inter-day precisions were below 10.2% and the accuracy was between -2.3% and +2.8%. The validated LC/MS/MS method was applied to a pharmacokinetic study in which healthy Chinese volunteers each received a single oral dose of 5 mg donepezil hydrochloride. The non-compartmental pharmacokinetic model was used to fit the donepezil plasma concentration-time curve. Maximum plasma concentration was 12.3 +/- 2.73 ng/mL which occurred at 3.50 +/- 1.61 h post-dosing. The apparent elimination half-life and the area under the curve were, respectively, 60.86 +/- 12.05 h and 609.3 +/- 122.2 ng h/mL. LC/MS/MS is a rapid, sensitive and specific method for determining donepezil in human plasma samples. Copyright (c) 2006 John Wiley & Sons, Ltd.