4.5 Article

GRadient echo Acquisition for Superparamagnetic particles with positive contrast (GRASP): Sequence characterization in membrane and glass superparamagnetic iron oxide phantoms at 1.5T and 3T

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 55, 期 1, 页码 126-135

出版社

WILEY
DOI: 10.1002/mrm.20739

关键词

iron oxide; gradient echo imaging; susceptibility; T-2(*); positive contrast MRI

资金

  1. NHLBI NIH HHS [HL78667, R01 HL71021] Funding Source: Medline
  2. NATIONAL HEART, LUNG, AND BLOOD INSTITUTE [R01HL078667, R01HL071021] Funding Source: NIH RePORTER

向作者/读者索取更多资源

Iron oxides are used for cell trafficking and identification of macrophages in plaque using MRI. Due to the negative contrast, differentiation between signal loss caused by iron and native low signal in tissue may be problematic. It is, therefore, preferable to achieve positive contrast. The purpose of this study was to test the efficacy of a new MRI sequence GRASP (GRe Acquisition for Superparamagnetic Particles) to generate a positive signal in phantoms containing iron. Membrane phantoms were constructed containing Ferumoxide at 7 concentrations. Standard GRE sequences were modified with user controlled z-gradient rephasing (+/- 100%). CNR values were determined as a function of echo time (TE) and % rephasing at 1.5T and 3T. T-2(star) values were determined using multiple double-echo GRE. The GRASP sequence generated positive signal enhancement in phantoms containing iron. For all rephasing values <= 30%, positive contrast was observed. The CNR generated at 1.5T was greater than the values at 3T for all concentrations tested. Correlation between CNR at 0% and 100% rephasing was observed at 1.5T(R = 0.84). Additionally, correlation between field change across the volume and CNR was observed. In conclusion, GRASP sequences may be used to generate positive signal enhancement in the presence of iron using MRI.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据