The determination of angiotensin-converting enzyme (ACE) activity represents a useful tool in the study of different health pathologies, such as hypertension. This protocol describes a fluorescent assay for measuring ACE activity in vitro with high precision and sensitivity. The method relies on the ability of ACE to hydrolyse the internally quenched fluorescent substrate o-aminobenzoylglycyl-p-nitro-L-phenylalanyl-L-proline. The generation of the fluorescent product o-aminobenzoylglycine can be continuously monitored, preferably using a microtiter-plate fluorometer, though the use of a conventional cuvette fluorometer would also be possible. The method has important advantages with respect to other assays, because it involves only a one-step reagent, is easy to carry out and allows the analysis of an elevated number of samples in shorter times. It can be completed in one and a half hours. In addition, the fact that all reagents are commercially available allows the rapid introduction of the assay into the laboratory.