4.6 Article

Determination of heavy metals in water samples using dual-cloud point extraction coupled with inductively coupled plasma mass spectrometry

期刊

ANALYTICAL METHODS
卷 7, 期 16, 页码 6732-6739

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c5ay00801h

关键词

-

资金

  1. National Water Pollution Control and Management Technology Major Projects [2012ZX07307-001]

向作者/读者索取更多资源

A dual-cloud point extraction (d-CPE) technique combined with inductively coupled plasma mass spectrometry (ICP-MS) was used for preconcentration and determination of Cr, Ga, Ag, Cd, Mn, Fe, In, Cu, Ni, Co, Pb and Zn in real water samples. An orthogonal array design (OAD) with the OA(25) (5(6)) matrix was employed to study the effects of different parameters on the extraction efficiency. Under the best experimental conditions (concentration of complexing reagent 8-hydroxyquinoline, 0.25 mmol L-1; pH, 6.5; equilibrium temperature, 60 degrees C; surfactant Triton X-114, 0.1%; equilibrium time, 30 min and without salt addition), the enhancement factor ranged from 9.85 to 35.98. The linear dynamic range (LDR) was 1-800 mg L-1 with correlation coefficients of 0.9974-0.9999 and the limits of detection (LODs) were between 0.012 and 0.36 mg L-1. Relative standard deviations (RSDs, C = 10 mg L-1 and n = 6) were 1.29%-5.56%. Finally, the developed method was successfully applied to extraction and determination of the mentioned metal ions in real water samples and satisfactory results were obtained.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据