Influence of display quality on radiologists' performance in the detection of lung nodules on radiographs

期刊

BRITISH JOURNAL OF RADIOLOGY
卷 80, 期 957, 页码 738-743

出版社

BRITISH INST RADIOLOGY
DOI: 10.1259/bjr/48049509

关键词

-

向作者/读者索取更多资源

The purpose of this study was to evaluate the influence of display quality on radiologists' performance in the detection of lung nodules. Display systems with various technical properties were considered based on their general availability in a radiology department. Their quality was assessed by physical tests. Multireader-multicase receiver operating characteristic (ROC) analysis was used to evaluate observer performance. The area under the curve (Az was used as a metric for detectability of simulated lung nodules with diameters of 5 mm and 10 mm, and peak contrast values ranging from 0.1 (subtle) to 0.4 (evident) that were digitally superimposed on normal chest radiographs. Three experienced radiologists interpreted a batch of 60 radiographs on five different display systems; four monitors (two liquid crystal display (LCD) and two cathode ray tube (CRT) monitors) and one printed hardcopy. The physical tests showed superior performance of the two LCD monitors. ROC analysis resulted in the following Az scores: LCD-5MP Az=0.78, hardcopy Az=0.77, LCDc-2MP Az=0.75, CRT-5MP Az=0.72 and CRTc-1MP Az=0.71. Difference in Az scores between the LCD-5MP monitor and both the CRT-5MP (p=0.04) and CRTc-1MP (p=0.01) monitors was significant. The primary class CRT-5MP monitor that showed reduced observer performance failed to comply with physical acceptance requirements. Luminance response was particularly observed to be insufficient. The results indicate that a quality assurance program has the potential to detect non-optimised display systems that could otherwise result in reduced observer performance.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据