4.6 Article

The performance of a nitrogen-containing tetrahedral amorphous carbon electrode in flow injection analysis with amperometric detection

期刊

ANALYTICAL METHODS
卷 7, 期 11, 页码 4481-4485

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c5ay01017a

关键词

-

资金

  1. Army Research Office [W911NF-12-R-0011]
  2. National Science Foundation [CHE-0911383]
  3. ARO
  4. NSF

向作者/读者索取更多资源

The performance of a nitrogen-incorporated tetrahedral amorphous (ta-C:N) carbon thin-film electrode was evaluated using flow injection analysis with amperometric detection. The response stabilization time, and background current and noise were evaluated as a function of the applied potential. The analytical detection figures of merit for norepinephrine in 0.1 M phosphate buffer (pH 7.2) were also determined for the ta-C: N electrode and compared with values obtained at a boron-doped nanocrystalline diamond film electrode. The ta-C: N electrode exhibited more rapid stabilization times, and lower background current and noise at all potentials from 0.1 to 0.9 V vs. Ag/AgCl. The response precision, sensitivity, linear dynamic range and limit of detection were as good or superior for the ta-C:N electrode. The results portend the analytical utility of this new carbon electrode material.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据