Methods for high-precision C-14 AMS measurement of atmospheric CO2 at LLNL

期刊

RADIOCARBON
卷 49, 期 2, 页码 349-356

出版社

UNIV ARIZONA DEPT GEOSCIENCES
DOI: 10.1017/S0033822200042284

关键词

-

向作者/读者索取更多资源

Development of radiocarbon analysis with precision better than 2%omicron has the potential to expand the utility of (CO2)-C-14 measurements for carbon cycle investigations as atmospheric gradients currently approach the typical measurement precision of 2-5%omicron. The accelerator mass spectrometer at Lawrence Livermore National Laboratory (LLNL) produces high and stable beam currents that enable efficient acquisition times for large numbers of C-14 counts. One million C-14 atoms can be detected in approximately 25 min, suggesting that near 1%omicron counting precision is economically feasible at LLNL. The overall uncertainty in measured values is ultimately determined by the variation between measured ratios in several sputtering periods of the same sample and by the reproducibility of replicate samples. Experiments on the collection of 1 million counts on replicate samples of CO2 extracted from a whole air cylinder show a standard deviation of 1.7%omicron in 36 samples measured over several wheels. This precision may be limited by the reproducibility of oxalic acid I standard samples, which is considerably poorer. We outline the procedures for high-precision sample handling and analysis that have enabled reproducibility in the cylinder extraction samples at the <2%omicron level and describe future directions to continue increasing measurement precision at LLNL.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据