JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE
卷 9, 期 4, 页码 673-679出版社
BMC
DOI: 10.1080/10976640601138805
关键词
steady state free precession; fast low angle shot; 3 Tesla; left atrium; biplane are-length method
资金
- British Heart Foundation [RG/07/004/22659, PS/02/002/14893] Funding Source: Medline
- Medical Research Council [G0400444] Funding Source: Medline
- Wellcome Trust Funding Source: Medline
- MRC [G0400444] Funding Source: UKRI
Purpose: To investigate left atrial volumes and function and their variability in healthy volunteers using steady state free precession (SSFP) and fast low angle shot (FLASH) sequences at both 1.5 and 3 T using both the short-axis and biplane area-length methods. Materials and Methods: Ten healthy volunteers underwent CMR at both 1.5 and 3 Tesla. The biplane area-length method utilized volumes from the horizontal and vertical long axis images. Results: There were no significant differences between left atrial short-axis volumes or function between 1.5 and 3 T assessed using either FLASH or SSFP sequences. The biplane area-length method underestimated maximal left atrial volume using FLASH by 12 mL at 3 T (18%) and by 10 mL (14%) at 1.5 T (p = 0.003 and p = 0.05 respectively). Variability was larger for left atrial measurements using the biplane area-length method. Conclusion: Field strength had no effect on left atrial volume and function assessment using either FLASH or SSFP. The use of the short-axis method for the acquisition of left atrial parameters is more reproducible than the biplane area-length for serial measurements.