Evaluation of oxygenators and centrifugal pumps for long-term pediatric extracorporeal membrane oxygenation

期刊

PERFUSION-UK
卷 22, 期 5, 页码 323-326

出版社

SAGE PUBLICATIONS LTD
DOI: 10.1177/0267659107086270

关键词

-

向作者/读者索取更多资源

Objective: Two extracorporeal membrane oxygenation ( ECMO) circuits for children under 10 kg were evaluated and compared for plasma leakage, hemolysis, blood transfusions, and durability. Methods: Group A ( n = 20) was supported by ECMO circuits with the Minimax oxygenator and the Biomedicus centrifugal pump. Group B ( n = 10) was supported by ECMO circuits with the Lilliput 2 ECMO oxygenator and the Rotaflow centrifugal pump. Results: ECMO circuit durability, as measured by oxygenator lifespan, was significantly better in Group B than in Group A ( p = 0.04). There was significantly lower hemolysis, measured by plasma free hemoglobin, in Group B ( p = 0.019), and patients in Group B had significantly less need for antithrombin III transfusion ( p = 0.004). No plasma leakage was observed in Group B oxygenators, but plasma leakage was observed in all Group A oxygenators. Conclusion: The combination of a Rotaflow centrifugal pump and Lilliput 2 ECMO oxygenator in pediatric ECMO circuits improved durability and reduced circuit-induced hemolysis. This improvement may be due to the low priming volume, the oxygenator's plasma leakage resistance, the suspended rotor of the centrifugal pump, or a combination of these factors.