4.7 Article

Bead-based immunoassays using a micro-chip flow cytometer

期刊

LAB ON A CHIP
卷 7, 期 8, 页码 1048-1056

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b707507n

关键词

-

资金

  1. Engineering and Physical Sciences Research Council [GR/S23513/01] Funding Source: researchfish

向作者/读者索取更多资源

A microfabricated flow cytometer has been developed for the analysis of micron-sized polymer beads onto which fluorescently labelled proteins have been immobilised. Fluorescence measurements were made on the beads as they flowed through the chip. Binding of antibodies to surface-immobilised antigens was quantitatively assayed using the device. Particles were focused through a detection zone in the centre of the flow channel using negative dielectrophoresis. Impedance measurements of the particles (at 703 kHz) were used to determine particle size and to trigger capture of the fluorescence signal. Antibody binding was measured by fluorescence at single and dual excitation wavelengths (532 nm and 633 nm). Fluorescence compensation techniques were implemented to correct for spectral overspill between optical detection channels. The data from the microfabricated flow cytometer was shown to be comparable to that of a commercial flow cytometer (BD-FACSAria).