4.7 Article

Pumping fluids in microfluidic systems using the elastic deformation of poly(dimethylsiloxane)

期刊

LAB ON A CHIP
卷 7, 期 12, 页码 1832-1836

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/b714664g

关键词

-

资金

  1. NIEHS NIH HHS [ES16665] Funding Source: Medline
  2. NATIONAL INSTITUTE OF ENVIRONMENTAL HEALTH SCIENCES [R01ES016665] Funding Source: NIH RePORTER

向作者/读者索取更多资源

This paper demonstrates a methodology for storing and pumping fluids that provide a useful capability for microfluidic devices. It uses microfluidic screw valves to isolate fluids in poly(dimethylsiloxane) (PDMS) microcompartments, in which the pressure of the liquid is stored in the elastic deformation of the walls and ceiling of the compartments. Fluids can be stored under pressure in these structures for months. When the valves are opened, the walls and ceiling push the fluid out of the compartments into microfluidic channels. The system has five useful characteristics: (i) it is made using soft lithographic techniques; (ii) it allows multiple reagents to be preloaded in devices and stored under pressure without any additional user intervention; (iii) it makes it possible to meter out fluids in devices, and to control rates of flow of fluids; (iv) it prevents the user from exposure to potentially toxic reagents; and (v) it is hand-operated and does not require additional equipment or resources.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据