4.7 Article Proceedings Paper

Flow injection on-line hydrophobic sorbent extraction for flame atomic absorption spectrometric determination of cadmium in water samples

期刊

MICROCHIMICA ACTA
卷 160, 期 4, 页码 455-460

出版社

SPRINGER WIEN
DOI: 10.1007/s00604-007-0775-2

关键词

cadmium; flow injection; solid phase extraction; atomic absorption spectrometry; poly-chlorotrifluoroethylene

向作者/读者索取更多资源

A flow injection system was developed for on-line sorbent extraction preconcentration and flame atomic absorption spectrometric determination of cadmium in natural water samples. The non-charged cadmium complex with diethyl-dithiophosphate (DDPA) was formed on-line in 0.1 mol L-1 HNO3 and retained on the hydrophobic poly-chlorotrifluoroethylene (PCTFE) sorbent material. The adsorbed complex was eluted with isobutyl methyl ketone (IBMK) and injected directly into the nebulizer via a flow compensation unit. All major chemical and flow parameters affecting the complex formation adsorption and elution as well as interference were studied and optimized. By processing 2.4 mL of sample, the enhancement factor was 39 and the sampling frequency was 50 h(-1). For 30s preconcentration time the detection limit was 0.3 mu g L-1 and the relative standard deviation at 5.0 mu g L-1 Cd concentration level was 2.9%. The calibration curve was linear in the range 0.8-40.0 mu g L-1. The accuracy of the method was estimated by analyzing a certified reference material NIST-CRM 1643d (Trace elements in water). Good recoveries were obtained for spiked natural-water and waste-water samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据