期刊
BUNSEKI KAGAKU
卷 59, 期 10, 页码 903-910出版社
JAPAN SOC ANALYTICAL CHEMISTRY
DOI: 10.2116/bunsekikagaku.59.903
关键词
ABS resin disk CRMs for Cd, Cr, Hg and Pb analysis; RoHS directive; ED-XRF analysis; stability; XRF sensitivity; homogeneity
ABS resin disk certified reference materials (CRMs) for heavy-metal analysis (NMIJ CRM 8105-a, NMIJ CRM. 8106-a, NMIJ CRM 8115-a and NMIJ CRM 8116-a) were evaluated using an energy dispersive X-ray fluoresence (ED-XRF) analysis. The homogeneities of elements for both among-disks and within-disk were evaluated by ED-XRF analysis without any sample pre-treatment, which were similar to those evaluated by ICP-MS analysis after a sample digenstion procedure: The normalized XRF sensitivities for Cd, Cr and Pb in different ABS resin disk CRMs were compared, and the differences among them for those ABS resin disks that have similar matrices were observed. Moreover, Hg in those ABS resin disk CRMs was stable for long-term X-ray irradiation during ED-XRF analysis.