4.6 Article

Source localization of auditory evoked potentials after cochlear implantation

期刊

PSYCHOPHYSIOLOGY
卷 45, 期 1, 页码 20-24

出版社

WILEY
DOI: 10.1111/j.1469-8986.2007.00610.x

关键词

auditory evoked potential; cochlear implant; independent component analysis; dipole source analysis; N1

向作者/读者索取更多资源

Little is known about how the auditory cortex adapts to artificial input as provided by a cochlear implant (CI). We report the case of a 71-year-old profoundly deaf man, who has successfully used a unilateral Cl for 4 years. Independent component analysis (ICA) of 61-channel EEG recordings could separate CI-related artifacts from auditory-evoked potentials (AEPs), even though it was the perfectly time-locked CI stimulation that caused the AEPs. AEP dipole source localization revealed contralaterally larger amplitudes in the P1-N1 range, similar to normal hearing individuals. In contrast to normal hearing individuals, the man with the CI showed a 20-ms shorter N1 latency ipsilaterally. We conclude that ICA allows the detailed study of AEPs in CI users.