期刊
ATMOSPHERIC MEASUREMENT TECHNIQUES
卷 2, 期 2, 页码 363-378出版社
COPERNICUS GESELLSCHAFT MBH
DOI: 10.5194/amt-2-363-2009
关键词
-
资金
- EUSAAR [RII3-CT-2006-026140]
- EC
- ESF
- Swiss National Science Foundation
- French National Research Agency (ANR)
We report on an intercomparison of six different hygroscopicity tandem differential mobility analysers (HTDMAs). These HTDMAs are used worldwide in laboratory experiments and field campaigns to measure the water uptake of aerosol particles and have never been intercompared. After an investigation of the different design of the instruments with their advantages and inconveniencies, the methods for calibration, validation and data analysis are presented. Measurements of nebulised ammonium sulphate as well as of secondary organic aerosol generated from a smog chamber were performed. Agreement and discrepancies between the instruments and to the theory are discussed, and final recommendations for a standard instrument are given, as a benchmark for laboratory or field experiments to ensure a high quality of HTDMA data.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐