Validation of Datascope Accutorr PlusTM using British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocol guidelines

期刊

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jsha.2010.03.001

关键词

AAMI; BHS; BP devices; Datascope Acutorr plus; Validation

向作者/读者索取更多资源

Objective: To assure the accuracy and reliability of blood pressure measurement by noninvasive blood pressure monitor using Datascope Accutorr Plus (TM) (Paramus, NJ, USA) against mercury manometer, among adult male participants. Method: Eighty participants from a family physician's office at a teaching hospital were recruited. One hundred and sixty measurements of blood pressure were performed according to BHS technique protocol. Statistical analysis: Descriptive analysis was done according to the AAMI and BHS protocol guidelines. The limits of agreement between the device and the standard were plotted using the method of Bland and Altman plot. Results: The mean difference + SD between the Datascope Acutorr Plus (TM) and observer was 2.7 +/- 5.2 mm Hg and 1.5 +/- 3.26 mm Hg for systolic and diastolic blood pressure, respectively. Datascope Acutorr Plus (TM) obtained A/A grading for both systolic and diastolic blood pressure. Conclusion: Datascope Accutorr Plus (TM) (Paramus, NJ, USA) satisfies BHS and AAMI validation protocols for both systolic and diastolic BP and may be recommended for everyday use for BP monitoring at home and in clinical use for adult population. (C) 2010 King Saud University. All rights reserved.