Beam hardening artifacts in micro-computed tomography scanning can be reduced by X-ray beam filtration and the resulting images can be used to accurately measure BMD

期刊

BONE
卷 45, 期 6, 页码 1104-1116

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.bone.2009.07.078

关键词

Micro-computed tomography; Beam hardening; Bone mineral density; Bone densitometry; Ash fraction

资金

  1. National Science Foundation
  2. NIH [P30-AR46024, T90-DK070071]
  3. University of Michigan Medical School

向作者/读者索取更多资源

Bone mineral density (BMD) measurements are critical in many research studies investigating skeletal integrity. For pre-clinical research, micro-computed tomography (mu CT) has become an essential tool in these studies. However, the ability to measure the BMD directly from mu CT images can be biased by artifacts, such as beam hardening, in the image. This three-part study was designed to understand how the image acquisition process can affect the resulting BMD measurements and to verify that the BMD measurements are accurate. In the first part of this study, the effect of beam hardening-induced cupping artifacts on BMD measurements was examined. In the second part of this study, the number of bones in the X-ray path and the sampling process during scanning was examined. In the third part of this study, mu CT-based BMD measurements were compared with ash weights to verify the accuracy of the measurements. The results indicate that beam hardening artifacts of up to 32.6% can occur in sample sizes of interest in studies investigating mineralized tissue and affect mineral density measurements. Beam filtration can be used to minimize these artifacts. The results also indicate that, for murine femora, the scan setup can impact densitometry measurements for both cortical and trabecular bone and morphologic measurements of trabecular bone. Last, when a scan setup that minimized all of these artifacts was used, the mu CT-based measurements correlated well with ash weight measurements (R-2=0.983 when air was excluded), indicating that mu CT can be an accurate tool for murine bone densitometry. (C) 2009 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据