Evaluation of coagulation stages of hemorrhaged swine: comparison of thromboelastography and rotational elastometry

期刊

BLOOD COAGULATION & FIBRINOLYSIS
卷 21, 期 1, 页码 20-27

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/MBC.0b013e32833113e9

关键词

coagulation; coagulopathy; swine; thromboelastography; thromboelastometry; trauma

资金

  1. BUMED [604771N.9737.001.A0315]
  2. MARCORPS [26623M.162M.A0607]

向作者/读者索取更多资源

Thromboelastography (TEG) or rotational thromboelastometry (ROTEM) assesses blood viscoelastic properties and clotting kinetics that can be measured by Haemoscope TEG and Pentapharm ROTEM devices using slightly different methodologies. These devices were compared by measuring blood samples associated with various degrees of coagulopathy. Blood samples, collected from swine undergoing three types of severe injury and resuscitation protocol resulting in normal, hypercoagulopathy, and hypocoagulopathy, were assessed with TEG or ROTEM before the surgical procedures, and after injury, fluid resuscitation, and simulated hospital phase. Standard clotting parameters were compared by Student's t-test at a significance of a P value less than 0.05. Regression analysis indicated a positive correlation between TEG and ROTEM for reaction time (R), clotting rate (K), and maximum amplitude (M,,) parameters. With samples of normal coagulation, R (440 +/- 136 vs. 391 +/- 73 s), K (99 +/- 39 vs. 81 +/- 20 s), and M-a (74 +/- 4 vs. 69 +/- 5 mm) were higher, whereas (alpha) (68 +/- 8 vs. 75 +/- 3 mm) was lower with TEG than ROTEM respectively; a P value is less than 0.05. The magnitude of changes from baseline in hypercoagulable or hypocoagulable samples due to level of injury was equivalent with TEG and ROTEM indicating comparable use of the instruments. However, when samples were extremely hypocoagulopathic due to resuscitation fluid, the TEG values could not be readily determined. Overall, TEG readings were higher than ROTEM readings; this disparity between the two instruments was attenuated with hypercoaguable samples. Both devices yielded similar information regarding the status of coagulation related to trauma. Because of operating characteristics, the same instrument should be used for monitoring the same patient or study. Blood Coagul Fibrinolysis 21:20-27 (C) 2010 Wolters Kluwer Health | Lippincott Williams & Wilkins.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据