期刊
BLOOD COAGULATION & FIBRINOLYSIS
卷 19, 期 6, 页码 569-576出版社
LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/MBC.0b013e3283070872
关键词
clot detection; coagulation; mechanical detection; optical detection
类别
资金
- Dade Behring Inc, Deerfield, Illinois, USA
Hemostasis instrumentation has rapidly advanced and laboratories are demanding fully automated coagulation systems. Two distinct technological families exist based on optical and mechanical clot detection methodologies. Until now, there have been no comprehensive studies to determine whether one methodology is superior to the other. In order to answer this question, we conducted a large clinical study performing standard coagulation testing on more than 2000 clinical samples randomly chosen from a high-volume laboratory in a tertiary care hospital. Results demonstrated that photo-optical clot detection and electromechanical detection systems were highly correlated (r-squared values >= 0.96 for all assays) Correlation between the two clot detection systems was maintained even when measuring turbid samples (r-squared values >= 0.98 for all assays).