4.4 Article

Determination of Albumin in Urine by a Quartz Crystal Microbalance Label-Free Assay

期刊

ANALYTICAL LETTERS
卷 50, 期 12, 页码 1912-1925

出版社

TAYLOR & FRANCIS INC
DOI: 10.1080/00032719.2016.1262393

关键词

Biosensor; immunosensors; microalbumin; piezoelectric device; quartz crystal microbalance

资金

  1. Faculty of Medical Technology, Mahidol University
  2. Commission on Higher Education (CHE)

向作者/读者索取更多资源

Albumin is a vital plasma protein in the control of osmotic pressure of human blood and is used as a screening marker for renal disease and Type I and Type II diabetic mellitus. A mass-sensitive immunosensor based on quartz crystal microbalance was used to determine microalbumin in urine on the antihuman albumin layer which was bonded to the protein A layer on the surface of a silver electrode. The analytical conditions were optimized for protein A and antihuman albumin on the electrode, and a calibration curve and the correlation between the microalbumin measured by the immunosensor and by a conventional method were evaluated. The results demonstrated that the optimal conditions for protein A coating on the silver electrode were 1mg/ml and 2h incubation at room temperature. The optimal conditions for antihuman albumin on the protein A-modified surface were 0.5mg/ml at 2h. The dose-response curve was linear from 0.0001 to 0.1mg/ml. The comparison for the determination of microalbumin in urine by this method and a conventional approach showed good correlation with R-2=0.999 (P<0.01). The receiver operating characteristic curve analysis showed 98.7% sensitivity, 100% specificity, and 1.00 for the area under the curve for the 0.02mg/ml cutoff value. This system was label free and easy to use with high sensitivity and specificity. Therefore, the system may be used in place of commercial methods in the clinical laboratory.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据