期刊
出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2015.05.010
关键词
Vortioxetine; Rat; UPLC-MS/MS; Plasma; Pharmacokinetics
In this work, a simple, sensitive and fast ultra performance liquid chromatography with tandem mass spectrometry (UPLC-MS/MS) method was developed and validated for the quantitative determination of vortioxetine in rat plasma. Plasma samples were processed with a protein precipitation. The separation was achieved by an Acquity UPLC BEH C18 column (2.1 mm x 50 mm, 1.7 mu m) column with a gradient mobile phase consisting of 0.1% formic acid in water and acetonitrile. Detection was carried out using positive-ion electrospray tandem mass spectrometry via multiple reaction monitoring (MRM). The validated method had an excellent linearity in the range of 0.05-20 ng/mL (R-2 > 0.997) with a lower limit of quantification (0.05 ng/mL). The extraction recovery was in the range of 78.3-88.4% for vortioxetine and 80.3% for carbamazepine (internal standard, IS). The intra- and inter-day precision was below 8.5% and accuracy was from -11.2% to 9.5%. No notable matrix effect and astaticism was observed for vortioxetine. The method has been successfully applied to a pharmacokinetic study of vortioxetine in rats for the first time, which provides the basis for the further development and application of vortioxetine. (C) 2015 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐