期刊
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
卷 1001, 期 -, 页码 124-130出版社
ELSEVIER
DOI: 10.1016/j.jchromb.2015.07.032
关键词
MALDI-MS; Ionic liquid matrix; Dispersive liquid-liquid microextraction; Phospholipids; Soybean
Ionic liquid matrix (ILM) is found to be a very versatile substance for analysis of broad range of organic molecules in matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) due to good solubility for a variety of analytes, formation of homogenous crystals and high vacuum stability of the matrix. In the present work, an ILM, cyno-4-hydroxycinnamic acid-butylamine (CHCAB) was employed in dispersive liquid-liquid microextraction (DLLME) as sample probe and matrix for extraction and ionization of phospholipids from food samples (soybean) prior to MALDI-MS analysis. With the employed technique, 8-125 fold improvement in signal intensity and limit of detection were achieved for the analysis of phospholipids. The best extraction efficiency of phospholipids in ILM-DLLME was obtained with 5 min extraction time in presence 30 mg/mL CHCAB and 1.2% NaCl using chloroform as an extracting solvent and methanol as a dispersing solvent. Further, the developed ILM-DLLME procedure has been successfully applied for the analysis of phospholipids in soybean samples in MALDI-MS. (C) 2015 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐