4.6 Article

Analysis of volatile phenols in alcoholic beverage by ethylene glycol-polydimethylsiloxane based stir bar sorptive extraction and gas chromatography-mass spectrometry

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1390, 期 -, 页码 22-27

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2015.02.064

关键词

Volatile phenols; Beer; Wine; SBSE-GC-MS; EG/PDMS phase

向作者/读者索取更多资源

An ethylene glycol (EG)/polydimethylsiloxane (PDMS) copolymer based stir bar sorptive extraction (SBSE)-GC-MS method was developed for the analysis of volatile phenols (4-ethylpheno1,4-vinylphenol, 4-ethylguaiacol, and 4-vinylguaiacol) in alcoholic beverages. The beverage samples were diluted with phosphate buffer (1 M, pH 7) and extracted with an EG/PDMS stir bar. Volatile phenols were thermally desorbed and analyzed by GC-MS. Parameters affecting extraction efficiency were studied including ionic strength, pH, extraction time, ethanol content and nonvolatile matrix. Good correlation coefficients with R-2 in the range of 0.994-0.999 were obtained for volatile phenol concentration of 5-500 mu g/L. Recovery for all phenols were from 95.7% to 104.4% in a beer matrix and 81.4% to 97.6% in a wine matrix. The method had a standard deviation less than 5.8% for all volatile phenols. The limit of quantification (LOQs) in beer samples was lower than 3 mu g/L. The method was further applied to analyze the concentrations of volatile phenols in beer, wine and other alcoholic beverage samples. (C) 2015 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据