期刊
JOURNAL OF CHROMATOGRAPHY A
卷 1386, 期 -, 页码 62-73出版社
ELSEVIER
DOI: 10.1016/j.chroma.2015.01.088
关键词
Liquid chromatography; High resolution mass spectrometry; Bivalve samples; Lipophilic marine biotoxins
资金
- Thermo Fisher Scientific Inc.
A new method for the analysis of lipophilic marine biotoxins (okadaic acid, dinophysistoxins, azaspiracids, pectenotoxins, yessotoxins, spirolids) in fresh and canned bivalves has been developed. A QuEChERS methodology is applied; i.e. the analytes are extracted with acetonitrile and clean-up of the extracts is performed by dispersive solid phase extraction with C-18. The extracts are analyzed by ultra-high performance liquid chromatography coupled to a hybrid quadrupole-Orbitrap mass spectrometer, operating in tandem mass spectrometry mode, with resolution set at 70,000 (m/z 200, FWHM). Separation of the analytes, which takes about 10min, is carried out in gradient elution mode with a BEH C-18 column and mobile phases based on 6.7 mM ammonia aqueous solution and acetonitrile mixtures. For each analyte the molecular ion and 1 or 2 product ions are acquired, with a mass accuracy better than 5 ppm. The quantification is performed using surrogate matrix matched standards, with eprinomectin as internal standard. The high-throughput method, which has been successfully validated, fulfills the requirements of European Union legislation, and has been implemented as a routine method in a public health laboratory. (c) 2015 Elsevier B.V. All rights reserved.