4.8 Article

Multiplexed detection of cardiac biomarkers in serum with nanowire arrays using readout ASIC

期刊

BIOSENSORS & BIOELECTRONICS
卷 35, 期 1, 页码 218-223

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2012.02.052

关键词

Silicon nanowire array; Biosensors; Cardiac biomarkers; Serum; Multiplexed detection; Application specific integrated circuit

资金

  1. Singapore Economic Development Board [COY-15-POC/1457-3]

向作者/读者索取更多资源

Early detection of cardiac biomarkers for diagnosis of heart attack is the key to saving lives. Conventional method of detection like the enzyme-linked immunosorbent assay (ELISA) is time consuming and low in sensitivity. Here, we present a label-free detection system consisting of an array of silicon nanowire sensors and an interface readout application specific integrated circuit (ASIC). This system provides a rapid solution that is highly sensitive and is able to perform direct simultaneous-multiplexed detection of cardiac biomarkers in serum. Nanowire sensor arrays were demonstrated to have the required selectivity and sensitivity to perform multiplexed detection of 100 fg/ml troponin T, creatine kinase MM, and creatine kinase MB in serum. A good correlation between measurements from a probe station and the readout ASIC was obtained. Our detection system is expected to address the existing limitations in cardiac health management that are currently imposed by the conventional testing platform, and opens up possibilities in the development of a miniaturized device for point-of-care diagnostic applications. (C) 2012 Elsevier BM. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据