4.8 Article

A membrane protein based biosensor: Use of a phosphate - H+ symporter membrane protein (Pho84) in the sensing of phosphate ions

期刊

BIOSENSORS & BIOELECTRONICS
卷 27, 期 1, 页码 58-63

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2011.06.024

关键词

Capacitive biosensor; Membrane protein; Pho84; Phosphate; Symporter

资金

  1. Swedish International Development Cooperation Agency (SIDA)

向作者/读者索取更多资源

A label free biosensor for direct detection of inorganic phosphate based on potential-step capacitance measurements has been developed. The high-affinity Pho84 plasma membrane phosphate/proton symporter of Saccharomyces cerevisiae was used as a sensing element. Heterologously expressed and purified Pho84 protein was immobilized on a self-assembled monolayer (SAM) on a capacitance electrode. Changes in capacitance were recorded upon exposure to phosphate compared to the control substance, phosphate analogue methylphosphonate. Hence, even without the explicit use of lipid membranes, the Pho84 membrane protein could retain its capacity of selective substrate binding, with a phosphate detection limit in the range of the apparent in vivo K-m. A linear increase in capacitance was monitored in the phosphate concentration range of 5-25 mu M. The analytical response of the capacitive biosensor is in agreement with that the transporter undergoes significant conformational changes upon exposure to inorganic phosphate, while exposure to the analogue only causes minor responses. (C) 2011 Published by Elsevier B.V.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据