4.8 Article

Direct electrochemical reduction of graphene oxide on ionic liquid doped screen-printed electrode and its electrochemical biosensing application

期刊

BIOSENSORS & BIOELECTRONICS
卷 28, 期 1, 页码 204-209

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2011.07.018

关键词

Electrochemically reduced graphene oxide; Screen-printed electrode; Ionic liquid; NADH; Glucose

资金

  1. Key Scientific Project of China [2009ZX08012-004B]

向作者/读者索取更多资源

A novel electrochemical biosensing platform using electrochemically reduced graphene oxide (ER-GNO) modified electrode was proposed. This modified electrode was prepared by one-step electrodeposition of the exfoliated GNO sheets onto the ionic liquid doped screen-printed electrode (IL-SPE). The resulting ER-GNO/IL-SPE brought new capabilities for electrochemical devices by combining the advantages of ER-GNO and disposable electrode. Two important biomolecules, nicotinamide adenine dinucleotide (NADH) and hydrogen peroxide (H2O2), were employed to study the electrochemical performance of the ER-GNO/IL-SPE, which exhibited more favorable electron transfer kinetics than the bare IL-SPE. On the basis of the greatly enhanced electrochemical reactivity of H2O2 at the developed electrode, ER-GNO and glucose oxidase constructed disposable biosensor showed better analytical performance for the glucose detection compared with the IL-SPE based biosensor. The linear range for the detection of glucose was from 5.0 mu M to 12.0 mM with a detection limit of 1.0 mu M. This work provides a useful avenue for implementing ER-GNO as a new generation of electrochemical transducer in disposable electrode, which could expand the scope of graphene constructed electrochemical biosensing devices and hold great promise for routine sensing applications. (C) 2011 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据