ZnO nanorods were grown on a silver-coated tip of a borosilicate glass capillary (0.7 mu m in tip diameter) and used as selective potentiometric sensor of intracellular free Mg2+. To functionalize the ZnO nanorods for selectivity of Mg2+, a polymeric membrane with Mg2+-selective ionophores were coated on the surface of the ZnO nanorods. These functionalized ZnO nanorods exhibited a Mg2+-dependent electrochemical potential difference versus an Ag/AgCl reference microelectrode within the concentration range from 500 nM to 100 mM. Two types of cells, human adipocytes and frog oocytes, were used for the intracellular Mg2+ measurements. The intracellular concentration of free Mg2+ in human adipocytes and frog oocytes were 0.4-0.5 and 0.8-0.9 mM, respectively. Such type of nanoelectrode device paves the way to enable analytical measurements in single living cells and to sense other bio-chemical species at the intracellular level. (C) 2010 Elsevier B.V. All rights reserved.