期刊
BIOSENSORS & BIOELECTRONICS
卷 25, 期 9, 页码 2089-2094出版社
ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2010.02.006
关键词
Screen-printed electrode; Polyaniline nanotube; Ionic liquid; Chronocoulometry; Hexaamineruthenium; DNA sensor
类别
资金
- Excellent Young Scientists Foundation of Shandong Province [R200805]
- University Doctoral Foundation of the Ministry of Education [200804260001]
- National Basic Research Program of China [2010CB732404]
A chronocoulometric DNA sensor was constructed based on doped screen-printed electrode (SPE). Polyaniline nanotubes (PANINTs) and ionic liquid (IL) [C(12)mim][PF6] were doped into the commercial graphite ink (Electrodag 423SS) for the printing of the SPE, and chitosan was covered on the electrode surface. A homogeneous, stable and highly conductive electrode surface was thus obtained. A chronocoulometric DNA sensor was established based on this SPE. The target DNA sequences were specifically recognized by the capture probes immobilized on the SPE surface and the reporter probes labeled on reporter Au nanoparticles, which also carried bio-bar-code DNA sequences. The signals were generated through chronocoulometric interrogation of [Ru(NH3)(6)](3+) ions that bound to the DNA strands (mainly bio-bar-code DNAs) via electrostatic interactions. The target DNA could be quantitatively detected in the range from 1.0 x 10(-16) M to 1.0 x 10(-14) M, and a detection limit of 8.0 x 10(-17) M was found. Selectivity against three-mismatched DNA was found in this sensor. (C) 2010 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐