期刊
BIOSENSORS & BIOELECTRONICS
卷 25, 期 7, 页码 1681-1687出版社
ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2009.12.007
关键词
Polyphenol oxidase; Catechol; Ionic liquid; Direct electropolymerization
类别
资金
- National Science Foundation of China [20873119]
A novel catechol biosensor was described based on the immobilization of polyphenol oxidase (PPO) into polyaniline (PANI), which was easily constructed by direct electropolymerization of aniline in a solution containing ionic liquid, 1-ethyl-3-methylimidazolium ethyl sulfate (EMIES). The developed biosensor for the detection of catechol has a linear range of 1.25-150 mu mol dm(-3). The maximum response current(I-max) and the Michaelis-Menten constant (k(m)') are 0.62 mu A and 146 mu mol dm(-3), respectively. The activation energy (E-a) of the PPO catalytic reaction is 31.1 kJ mol(-1) in the B-R buffer. The biosensor shows good reproducibility (a relative standard deviation of 3.1% was obtained) and remarkable long-term stability (it retains 75% of the original activity after four months). The effects of potential and pH on the response current of the biosensor are also described. (C) 2009 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐