期刊
BIOSENSORS & BIOELECTRONICS
卷 24, 期 7, 页码 2137-2142出版社
ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2008.11.016
关键词
QCM-D; E. coli O157:H7; Dissipation slope; Cysteamine; Self-assembled monolayer
类别
资金
- NSERC
- FQRNT
- FQRNT Centre for Biorecognition and Biosensors
- Canada Research Chairs Program
- Canada Foundation for Innovation
A biosensor for detection of viable Escherichia coli (E. coli) O157:H7 is developed using a quartz crystal microbalance with dissipation monitoring (QCM-D). The detection platform is based on the immobilization of affinity purified polyclonal antibodies onto gold-coated QCM-D quartz crystals via a cysteamine self-assembled monolayer. QCM-D measurements conducted over a broad range of bacterial cell concentrations show that the optimal biosensor response is the initial slope of the dissipation shift as a function of elapsed time (D-slope). A highly log-log linear response in the initial 0,1,p, is obtained for detection of E. coli O157:H7 over a wide range of cell concentrations from 3 x 10(5) to 1 x 10(9) cells/mL The prepared biosensor also exhibits a log-log linear working range from to 10(7) 10(9) cells/mL for E. coli K12 D21, a non-pathogenic model organism and further shows satisfactory selectivity using Bacillus subtilis. To our knowledge, this is the first study demonstrating the use of the initial D,I,p, as a sensor response when using QCM-D technology. (C) 2008 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐