期刊
BIOSENSORS & BIOELECTRONICS
卷 24, 期 6, 页码 1655-1660出版社
ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2008.08.040
关键词
Glucose sensor; Nanoscale nickel hydroxide; Blood samples; Ionic liquid; Ascorbic acid; Uric acid
类别
资金
- Third World Academy of Sciences Iran Chapter
- Shiraz University Research Council
The direct electrocatalytic oxidation of glucose in alkaline medium at nanoscale nickel hydroxide modified carbon ionic liquid electrode (CILE) has been investigated. Enzyme free electro-oxidation of glucose have greatly been enhanced at nanoscale Ni(OH)(2) as a result of electrocatalytic effect of Ni+2/Ni+3 redox couple. The sensitivity to glucose was evaluated as 202 mu A mM(-1) cm(-2). From 50 mu M to 23 mM of glucose can be selectively measured using platelet-like Ni(OH)(2) nanoscale modified CILE with a detection limit of 6 mu M (S/N = 3). The nanoscale nickel hydroxide modified electrode is relatively insensitive to electroactive interfering species such as ascorbic acid (AA), and uric acid (UA) which are commonly found in blood samples. Long-term stability, high sensitivity and selectivity as well as good reproducibility and high resistivity towards electrode fouling resulted in an ideal inexpensive amperometric glucose biosensor applicable for complex matrices. (C) 2008 Elsevier B.V. All rights reserved.