期刊
BIOSENSORS & BIOELECTRONICS
卷 24, 期 5, 页码 1336-1340出版社
ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2008.07.072
关键词
Amperometric biosensor; Urine glucose; Restricted permeable membrane; Clinical application
The highly sensitive urine glucose meter based on amperometric glucose sensor was developed and commercialized. It shows remarkable performances of wide measurement range in 0-2000 mg dl(-1), rapid response time as 6 s and robustness against influence by interferents like ascorbic acid or acetaminophen. Correlation between the developed urine glucose meter and commercialized clinical-use urine glucose analyzer showed excellent linear relationship. The monitoring of postmeal blood glucose levels by assess of urine glucose of actual subjects was performed with the developed urine glucose meter. The experimental results suggest the urine glucose level 120 min following the meal should be the appropriate index for diabetes or impaired glucose tolerance to control blood glucose level. The new portable meter was developed, and is expected for flexible use at places other than home or office. Crown Copyright (C) 2008 Published by Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐