4.8 Article

Novel highly-performing immunosensor-based strategy for ochratoxin A detection in wine samples

期刊

BIOSENSORS & BIOELECTRONICS
卷 23, 期 7, 页码 995-1002

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2007.10.002

关键词

electrochemical immunosensors; indirect competitive ELISA; ochratoxin A; wine

向作者/读者索取更多资源

The increasing concern about ochratoxin A (OTA) contamination of different food and feedstuffs demands high-performing detection techniques for quality assessment. Two indirect competitive enzyme-linked immunosorbent assay (ELISA) strategies were investigated for the development of OTA electrochemical immunosensors based on different OTA immobilisation procedures. Immunosensors based on avidin/biotin-OTA showed enhanced performance characteristics compared to those based on the adsorption of bovine serum albumin (BSA)-OTA conjugate. Performance of polyclonal (PAb) and monoclonal (MAb) antibodies against OTA was compared, showing at least one-order of magnitude lower IC50 values when working with MAb. Alkaline phosphatase (ALP)- and horseradish peroxidase (HRP)-labelled secondary antibodies were evaluated. Both conjugates led to similar results when working with OTA standard solutions in buffer. However, whereas electroactive interferences present in spiked wine samples did not affect HRP-labelled immunosensors (4% slope deviation), they were likely oxidised at 0.225 V versus Ag/AgCl, the working potential for ALP-labelled immunosensors (25% slope deviation). Considering 80% of antibody binding as the limit of detection, values of 0.7 and 0.3 ng/mL for HRP- and ALP-labelled immunosensors respectively, validate these immunosensors as useful screening tools to assess OTA levels in wine. (C) 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据