Use of oscillometric devices in atrial fibrillation: a comparison of three devices and invasive blood pressure measurement

期刊

BLOOD PRESSURE
卷 27, 期 1, 页码 48-55

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/08037051.2017.1383852

关键词

Blood pressure; atrial fibrillation; oscillometric device; hypertension; limits of agreement; invasive blood pressure

资金

  1. Lausanne University Hospital

向作者/读者索取更多资源

Background: The use of automated (oscillometric) blood pressure (BP) devices is not validated in atrial fibrillation (AF) patients. Objectives: To assess the reliability of three oscillometric BP devices, and the agreement with invasive arterial blood pressure(IBP) in AF patients. Methods: 48 AF patients with randomized sequences of 10 consecutive BP measurements with two pairs of devices: (1) OmronR7 (TM)(wrist) and OmronHEM907 (TM)(arm); (2) OmronR7 (TM) and Microlife WatchBPhome(arm). Reliability and agreement of each device were assessed by the intra-class correlation coefficient (ICC) for the continuous BP measurements and Bland & Altman methodology, respectively. In 10 additional AF patients, 10 consecutive measurements with IBP and OmronHEM907 (TM), and IBP and Microlife WatchBPhome were performed. Results: The OmronR7 (TM) was not able to obtain any BP Readings. Arm devices presented better ICC for systolicBP(SBP) than for diastolicBP(DBP) (Omron HEM907 (TM):0.94 [0.90; 0.97] vs. 0.77 [0.67; 0.89]; Microlife WatchBPhome: 0.92 [0.88; 0.96] vs. 0.79 [0.69; 0.89]). The correlation coefficient between Microlife WatchBPhome and IBP computed using the average of repeated measurements from two to ten measurements improved up to the third and remained stable afterwards. The agreement between IBP and SBP, and IBP and DBP, was moderate as illustrated by a wide limit of agreement [-24; 26](SBP) and [-15; 17](DBP) for Microlife WatchBPHome, respectively and [-30; 13](SBP) and [-7; 15](DBP) for OmronHEM907. Conclusions: BP measurement using the two arm oscillometric devices achieved a high reliability for SBP. The agreement between IBP and arm devices was low but using the average of three consecutive measurements improved the results substantially.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据