期刊
BIOMEDICAL ENGINEERING LETTERS
卷 8, 期 4, 页码 399-404出版社
SPRINGERNATURE
DOI: 10.1007/s13534-018-0085-0
关键词
Digital automatic blood pressure; Validation; Clinical use; Self-management
资金
- SK Telecom Co. Ltd
- Technology Innovation Program - Ministry of Trade, Industry and Energy (MOTIE) of Korea [10049785]
The purpose of this study was to evaluate the accuracy of a mobile wireless digital automatic blood pressure monitor for clinical use and mobile health (mHealth). In this study, a manual sphygmomanometer and a digital blood pressure monitor were tested in 100 participants in a repetitive and sequential manner to measure blood pressure. The guidelines for measurement used the Korea Food & Drug Administration protocol, which reflects international standards, such as the American National Standard Institution/Association for the Advancement of Medical Instrumentation SP 10: 1992 and the British Hypertension Society protocol. Measurements were generally consistent across observers according to the measured mean +/- SD, which ranged in 0.1 +/- 2.6 mmHg for systolic blood pressure (SBP) and 0.5 +/- 2.2 mmHg for diastolic blood pressure (DBP). For the device and the observer, the difference in average blood pressure (mean +/- SD) was 2.3 +/- 4.7 mmHg for SBP and 2.0 +/- 4.2 mmHg for DBP. The SBP and DBP measured in this study showed accurate measurements that satisfied all criteria, including an average difference that did not exceed 5 mmHg and a standard deviation that did not exceed 8 mmHg. The mobile wireless digital blood pressure monitor has the potential for clinical use and managing one's own health.