4.6 Article

A Label-Free Fluorescent AND Logic Gate Aptasensor for Sensitive ATP Detection

期刊

SENSORS
卷 18, 期 10, 页码 -

出版社

MDPI
DOI: 10.3390/s18103281

关键词

ATP detection; label-free fluorescence; enzyme-free; graphene oxide; logic gate

资金

  1. Natural Science Foundation of Hainan Province [518QN215, 2018CXTD332]
  2. National Nature Science Foundation of China [21763009]
  3. Hainan Province graduate student innovative research projects [Hys2018-59, Hys2018-60]

向作者/读者索取更多资源

In this study, a label-free fluorescent, enzyme-free, simple, highly sensitive AND logic gate aptasensor was developed for the detection of adenosine triphosphate (ATP). Double-stranded deoxyribonucleic acid (DNA) with cohesive ends was attached to graphene oxide (GO) to form an aptasensor probe. ATP and single-stranded DNA were used as input signals. Fluorescence intensity of PicoGreen dye was used as an output signal. The biosensor-related performances, including the logic gate construction, reaction time, linearity, sensitivity, and specificity, were investigated and the results showed that an AND logic gate was successfully constructed. The ATP detection range was found to be 20 to 400 nM (R-2 = 0.9943) with limit of detection (LOD) of 142.6 pM, and the sensitivity range was 1.846 x 10(6) to 2.988 x 10(6) M-1. This method for the detection of ATP has the characteristics of being simple, low cost, and highly sensitive.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据