4.8 Article

An instrument for simultaneous EQCM impedance and SECM measurements

期刊

ANALYTICAL CHEMISTRY
卷 72, 期 2, 页码 349-356

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac990796o

关键词

-

向作者/读者索取更多资源

A novel combination of an electrochemical quartz crystal microbalance (EQCM) and a scanning electrochemical microscope (SECM) has been built. Unlike conventional EQCMs, the instrument described here allows rapid in situ measurement of the modulus of the quartz crystal's transfer function. Data analysis in the complex plane for the Butterworth-Van Dyke (BVD) equivalent circuit yields the real and the imaginary components R (damping resistance) and XL (reactive inductance) of the crystal's electroacoustic impedance around its resonant frequency of 10 MHz, The influence of different tip shapes of an approaching microelectrode on the electroacoustic impedance of the quartz crystal was studied and found to be minimal for certain geometries, The capability of the EQCM/SECM instrument was tested in cyclic voltammetric plating/stripping experiments using a copper(I) chloride solution of high concentration in 1 M HCl, Four parameters, XL, R, the substrate, and the tip current, can be recorded simultaneously. Depletion layer effects were observed and could be corrected for to yield accurate current efficiencies for potentiodynamic and potentiostatic copper plating. The amperometric response of the SECM tip positioned closely to the substrate reflects the concentration changes of electroactive ions in the diffusion layer of the substrate electrode.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据