Determination of fetal heart rate from abdominal signals: evaluation of beat-to-beat accuracy in relation to the direct fetal electrocardiogram

期刊

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/bmt-2011-0130

关键词

abdominal signals; fetal electrocardiography; fetal heart rate; fetal monitoring; fetal scalp ECG

资金

  1. Polish National Science Center

向作者/读者索取更多资源

The main aim of our work was to assess the reliability of indirect abdominal electrocardiography as an alternative to the commonly used Doppler ultrasound monitoring technique. As a reference method, we used direct fetal electrocardiography. Direct and abdominal signals were acquired simultaneously, using dedicated instrumentation. The developed method of maternal signal suppression as well as fetal QRS complexes detection was presented. Recordings were collected during established labors, each consisted of four signals from the maternal abdomen and the reference signal acquired directly from the fetal head. After assessing the performance of the QRS detector, the accuracy of fetal heart rate measurement was evaluated. Additionally, to reduce the influence of inaccurately detected R-waves, some validation rules were proposed. The obtained results revealed that the indirect method is able to provide an accuracy sufficient for a reliable assessment of fetal heart rate variability. However, the method is very sensitive to recording conditions, influencing the quality of signals. Our investigations confirmed that abdominal electrocardiography, even in its current stage of development, offers an accuracy equal to or higher than an ultrasound method, at the same time providing some additional features.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据