Performance test of electron cyclotron resonance ion sources for the Hyogo Ion Beam Medical Center

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 71, 期 2, 页码 987-989

出版社

AMER INST PHYSICS
DOI: 10.1063/1.1150367

关键词

-

向作者/读者索取更多资源

Two electron cyclotron resonance (ECR) ion sources were manufactured for the accelerator facility at the Hyogo Ion Beam Medical Center. H-2(+), He2+, and C4+ were chosen as the accelerating ions because they have the highest charge to mass ratio among ion states which satisfy the required intensity and quality. The sources have the same structure as the 10 GHz ECR source at the Heavy Ion Medical Accelerator in Chiba except for a few improvements in the magnetic structure. Their performance was investigated at the Sumitomo Heavy Industries factory before shipment. The maximum intensity was 1500 mu A for H-2(+), 1320 mu A for He2+, and 580 mu A for C4+ at the end of the ion source beam transport line. These are several times higher than required. Sufficient performance was also observed in the flatness and long-term stability of the pulsed beams. These test results satisfy the requirements for medical use. (C) 2000 American Institute of Physics. [S0034-6748(00)53802-7].