4.3 Article

A soft and flexible biosensor using a phospholipid polymer for continuous glucose monitoring

期刊

BIOMEDICAL MICRODEVICES
卷 11, 期 4, 页码 837-842

出版社

SPRINGER
DOI: 10.1007/s10544-009-9300-1

关键词

Biosensor; Phospholipid polymer; Flexible; Glucose sensor

资金

  1. Japan Society for the promotion of Science (JSPS)
  2. MEXT

向作者/读者索取更多资源

A flexible biosensor using a phospholipid polymer to immobilization of glucose oxidase (GOD) was fabricated and tested. At first, an enzyme membrane formed by immobilizing GOD onto a porous polytetrafluoroethylene (PTFE) membrane using the phospholipid polymer (2-methacryloyloxyethyl phosphorylcholine (MPC) copolymerized with 2-ethylhexylmethacrylate (EHMA) : PMEH) was evaluated. According to the result of amperometric measurement, average density of GOD to be immobilized was optimized to 38.9 units cm(-2). Temperature and pH dependences were also investigated. Then, a flexible glucose sensor was fabricated by immobilizing GOD onto a flexible hydrogen peroxide electrode using PMEH. The flexible glucose sensor showed a linear relationship between output currents and glucose concentration in 0.05-1.00 mmol L-1, with a correlation coefficient of 0.999. The calibration range covered the normal tear glucose level of 0.14-0.23 mmol L-1. This indicates that the flexible biosensor is considered to be useful for monitoring of glucose in tear fluids.