期刊
BIOMEDICAL MICRODEVICES
卷 11, 期 4, 页码 837-842出版社
SPRINGER
DOI: 10.1007/s10544-009-9300-1
关键词
Biosensor; Phospholipid polymer; Flexible; Glucose sensor
资金
- Japan Society for the promotion of Science (JSPS)
- MEXT
A flexible biosensor using a phospholipid polymer to immobilization of glucose oxidase (GOD) was fabricated and tested. At first, an enzyme membrane formed by immobilizing GOD onto a porous polytetrafluoroethylene (PTFE) membrane using the phospholipid polymer (2-methacryloyloxyethyl phosphorylcholine (MPC) copolymerized with 2-ethylhexylmethacrylate (EHMA) : PMEH) was evaluated. According to the result of amperometric measurement, average density of GOD to be immobilized was optimized to 38.9 units cm(-2). Temperature and pH dependences were also investigated. Then, a flexible glucose sensor was fabricated by immobilizing GOD onto a flexible hydrogen peroxide electrode using PMEH. The flexible glucose sensor showed a linear relationship between output currents and glucose concentration in 0.05-1.00 mmol L-1, with a correlation coefficient of 0.999. The calibration range covered the normal tear glucose level of 0.14-0.23 mmol L-1. This indicates that the flexible biosensor is considered to be useful for monitoring of glucose in tear fluids.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐