4.2 Article

Loss of ELISA specificity due to biotinylation of monoclonal antibodies

期刊

JOURNAL OF IMMUNOLOGICAL METHODS
卷 235, 期 1-2, 页码 91-99

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0022-1759(99)00222-7

关键词

ELISA specificity; biotinylated antibodies; biotinylation ratio; surface plasmon resonance

向作者/读者索取更多资源

A significant degree of nonspecificity was found in ELISA determinations of soluble urokinase receptor (suPAR) in human blood plasma when biotinylated monoclonal antibodies (Mabs) were used for the detection layer. Surface plasmon resonance studies using both nonbiotinylated and biotinylated antibodies demonstrated that biotinylation reduced specific binding of the antibodies to their target antigen, suPAR. Furthermore, biotinylation produced a new interaction with unknown human plasma protein(s), unrelated to suPAR. Nonspecific interaction with plasma protein(s) was also observed after biotinylation of a Mab having no specific target antigen in human plasma and, in both cases, the level of nonspecific interaction was directly related to the degree of antibody biotinylation. These results reinforce earlier observations that biotinylation of antibodies can reduce the affinity of antibodies, but also indicate that, in addition, biotinylation can reduce the specificity of immunoassays for plasma proteins. (C) 2000 Elsevier Science B.V. All rights reserved.