4.7 Article

Amperometric enzyme electrode for the determination of hydrogen peroxide based on sol-gel/hydrogel composite film

期刊

ANALYTICA CHIMICA ACTA
卷 407, 期 1-2, 页码 111-118

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0003-2670(99)00778-3

关键词

biosensor; sol-gel; horseradish peroxidase; hydrogen peroxide; hydrogel; hybrid material

向作者/读者索取更多资源

A new type of organic-inorganic composite material was prepared by sol-gel method, and a peroxidase biosensor was fabricated by simply dropping sor-gel-peroxidase mixture onto glassy carbon electrode surface. The sol-gel composite film and enzyme membrane were characterized by Fourier-transform infrared (FT-IR) spectroscopy and EQCM, the electrochemical behavior of the biosensor was studied with potassium hexacyanoferrate(II) as a mediator, and the effects of pH and operating potential were explored for optimum analytical performance by using amperometric method. The response time of the biosensor was about 10 s; the linear range was up to 3.4 mM with a detection limit of 5 x 10(-7) M. The sensor also exhibited high sensitivity (15 mu A mM(-1)) and good long-term stability. In addition, the performance of the biosensor was investigated using flow injection analysis (FIA), and the determination of hydrogen peroxide in real samples was discussed. (C)2000 Elsevier Science B.V. All rights reserved.