Validation of a LC method for the determination of 5-aminosalicylic acid and its metabolite in plasma and urine

期刊

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/S0731-7085(99)00292-7

关键词

5-aminosalicylic acid; 5-amino-2-hydroxybenzoic acid; mesalazine; N-acetyl-5-aminosalicylic acid; high-performance liquid chromatography; validation

向作者/读者索取更多资源

The choice of a proper analytical method for the quantification of drugs and/or their metabolites in biological samples plays a significant role in the evaluation and interpretation of bioavailability, bioequivalence and pharmacokinetic data. The aim of this study was validation of a method for the identification and quantitative determination of 5-aminosalicylic acid (5-ASA) and its metabolite N-acetyl-5-aminosalicylic acid in human plasma and urine. According to previous studies on the disposition of 5-ASA (mesalazine) in a patient with inflammatory bowel diseases, we have developed a rapid, sensitive method for the determination of 5-ASA and its acetylated metabolite, N-acetyl-5-aminosalicylic acid (Ac-5-ASA). The advantage of this method is that it measures both compounds, and is more rapid, reproducible and credible than the previous studies [C. Fischer, K. Maier, U. Klotz, J. Chromatogr. Biomed. Appl. 225 (1981) 498-503; P.N. Shaw, A.L. Sivner, L. Aarons. J.B. Houston, J. Chromatogr. Biomed. Appl. 274 (1983) 393-397; B. Norlander, R. Gotthard, M. Strom, Aliment. Pharmacol. Ther. 3 (1989) 333-342; U. Klotz, G.L. Stracciari, Arzneim.-Forsch. Drug Res. II 43 (12) (1993) 1357-1359]. 5-ASA was quantitatively determined in human plasma and urine samples by liquid chromatography following prior derivatization to its acetylated metabolite (Ac-5-ASA). N-Acetyl-anthranilic acid was used as the internal standard. The detection was performed with a spectrofluorimetric detector, excitation at 311 nm, cut-off at 449 nm. The method was validated for the following parameters: linearity, recovery, sensitivity, precision, accuracy, selectivity and stability, limits of quantification and of detection. It showed good linearity (r(2) greater than or equal to 0,996) in the range 0.1 ng/ml to 8 mu g/ml using a Lichrospher 60 RP-select B column. The lower limit of detection was 20 ng/ml in plasma and urine. The within-run relative standard deviations (R.S.D.) were below 6.7% at all concentration levels and the between-run R.S.D. were below 25.4% at all concentration levels. (C) 2000 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据