4.4 Article

Development of a new UPLC-MSMS method for the determination of temozolomide in mice: application to plasma pharmacokinetics and brain distribution study

期刊

BIOMEDICAL CHROMATOGRAPHY
卷 27, 期 7, 页码 889-893

出版社

WILEY-BLACKWELL
DOI: 10.1002/bmc.2877

关键词

temozolomide; LC-MS; MS; pharmacokinetics; mice; brain distribution

向作者/读者索取更多资源

A sensitive and accurate liquid chromatography method with mass spectrometry detection was developed and validated for the quantification of temozolomide in mouse plasma and brain. Theophyllin was used as the internal standard. A single-step protein precipitation was used for plasma and brain sample preparation. The method was validated with respect to selectivity, extraction recovery, linearity, intra- and inter-day precision and accuracy, limit of quantification and stability. The method has a limit of quantification of 50ng/mL for temozolomide in plasma and 125ng/g in brain. This method was used successfully to perform brain and plasma pharmacokinetic studies of temozolomide in mice after intraperitoneal administration. Copyright (c) 2013 John Wiley & Sons, Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据