关键词
-
A compact device to derive the optical density of human macular pigment (MP) using heterochromatic flicker photometry is described. The validity of the system is assessed by measuring the optical density spectra of MP in 12 healthy subjects and comparing this with well-established previously published values. The mean spectral absorbance characteristics of MP across subjects corresponds well with accepted values. As reported in other studies, our measurements show a wide Variation of MP optical densities between individuals. In our technique within-subject variability is low; standard deviations are between 0.025 and 0.15 in most cases. The overall optical density of MP ranged from 0.08 to 0.84 with a mean of 0.496 and standard deviation of 0.257 at 460 nm. The stimulus size was 0.95 degrees. The unique feature of the technique is that it allows free viewing (not Maxwellian View) of the stimulus, it can be conducted easily and quickly and does not need frequent re-calibration. (C) 2000 The College of Optometrists. Published by Elsevier Science Ltd. All rights reserved.