A simple and reliable method was developed for determining blood concentrations of tramiprosate using reversed-phase HPLC with evaporative light scattering detection. The optimum evaporator and nebulizer temperatures for ELSD were set at 90 and 60 degrees C respectively. The method was linear for a concentration range of 10-1000 mu g/mL when 0.3 mL blood was used. The intra-assay precision ranged from 1.84 to 5.24%, and the coefficient of variation (CV) for a quality control sample at 10 mu g/mL was 5.24% with a bias of -9.50% from the target value. The inter-assay precision ranged from 2.58 to 5.96%, and the CV for a quality control sample at 10 mu g/mL was 5.96% with a bias of -7.50% from the target value. The method described here is suitable for management of tramiprosate in Alzheimer disease therapy. Copyright (C) 2010 John Wiley & Sons, Ltd.