4.8 Article

Selectivity of potentiometric ion sensors

期刊

ANALYTICAL CHEMISTRY
卷 72, 期 6, 页码 1127-1133

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac991146n

关键词

-

资金

  1. NIGMS NIH HHS [R01-GM58589] Funding Source: Medline

向作者/读者索取更多资源

Selectivities of solvent polymeric membrane ion-selective electrodes (ISEs) are quantitatively related to equilibria at the interface between the sample and the electrode membrane. However, only correctly determined selectivity coefficients allow accurate predictions of ISE responses to real-world samples. Moreover, they are also required for the optimization of ionophore structures and membrane compositions. Best suited for such purposes are potentiometric selectivity coefficients as defined already in the 1960s. This paper briefly reviews the basic relationships and focuses on possible biases in the determination of selectivity coefficients. The traditional methods to determine selectivity coefficients (separate solution method, fixed interference method) are still the same as those originally proposed by IUPAC in 1976. However, several precautions are needed to obtain meaningful data. For example, errors arise when the response to a weakly interfering ion is also influenced by the primary ion leaching from the membrane. Wrong selectivity coefficients may be also obtained when the interfering agent is highly preferred and the electrode shows counterion interference. Recent advances show how such pitfalls can be avoided. A detailed recipe to determine correct potentiometric selectivity coefficients unaffected by such biases is presented.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据